GC772x series of Gas/Liquid Chromatographs precisely separates the constituents present in an injected gas/liquid sample. Along with WinQCDS - Quazar Chromatography Data Acquisition-Analysis Software , it becomes really a very versatile, user-friendly and a powerful chemical analysis tool.
Take a look at the critical specifications of GC772x series of Gas/Liquid Chromatographs.
Oven temperature range 2o C above ambient to 450o C.
Temperature stability 0.03o C.
Minimum detectable limit less than 5 pg of carbon as ethylene using N2 carrier.
Linear dynamic range 107 .
WinQCDS 9.0 an advanced instrument Control and Data Analysis Software.
Versatile peak detection & editing tools.
Bunching, Savitsky, Golay & Adaptive Filtering for noise reduction.
Precise Area calculations algorithms.
Choice of split-less &/or split injectors.
Rapid auto cooling of oven without door opening.
Auto flame out protection.
State of the art electro-pnenumetic flow control.
EPC pressure stability 0.005 psi.
EPC flow stability 0.005 ml/min.
For more information and purchase-RELATED QUERRIES, please send us an enquiry here .
Oven
Operating Temperature Range
4°C above ambient to 450ºC
Temperature Control
PID type
Temperature Programming
7 ramps from keypad & as many as desired from PC
Skin temp.(at temp. 450ºC)
Less than (ambient + 20ºC)
Cooling rate (450ºC to 70ºC)
<9 minutes
Volume of the inner oven
15 L (23.5 X 24.5 X 26.0 cm³)
Max. Temp. gradient
+/-0.5ºC (over the column)
Temperature Stability
0.03ºC
Temp Set-point Resolution
1ºC
Ramp Resolution
0.1ºC/min
Ramp Rates(Oven)
Temp. range( ºC )
Ramp rates(ºC/Min )
50-70
>70
70-115
>50
115-175
>45
175-300
>40
300-400
>25
400-450
>15
Flame Ionisation Detector (FID)
Max. Operating Temperature
450ºC
Temp Set-point Resolution
1ºC
Minimum Detectable Limit
<5 pg of carbon/s as ethylene using N2 carrier
Sensitivity
0.008 Coulombs/gC
Linear Dynamic range
107
No. of settings for enhanced/attenuated sensitivity
2 (0.1 & 0.01µAFS)
Sampling rate
10/50Hz
Noise @ 10Hz & 10,000pA F.S.
<5µV(50fA)
Auto flame out protection
Yes
Injectors
Inlets available
Split / Splitless with purge
Max. Operating temperature
450ºC
Temperature Control
PID type
Temp Set-point Resolution
1ºC
Installation Requirements
Dimensions
79cmx50cmx51cm (BxDxH)
Power Requirement
220Vac/50Hz (stabilized) 3KVA max.
Operating Temperature
0ºC to 50ºC
Weight
48Kg
Electro-Pneumatic Flow
Modes of operation Packed Pressure Controlled Packed Flow Controlled Capillary Pressure Controlled Capillary Flow Controlled Without Split Capillary Flow Controlled With Split
Pressure stability
< 0.005 psi
Flow stability
< 0.005 ml/min.
Max. Pressure Control Point
100 psi
Max. Flow Control Point
100 ml/min.
Flow/Pressure Programming
Number of ramps & holds
As per user requirement
Number of lines Controlled
5 (packed) & 7 (capillary)
Computer Controlled Operation
Number of login levels
3
Monitoring Resolution
Flow
0.000001 ml/min.
Pressures
0.000001 psi
Temperatures
0.01o C
System Noise, Wander, Drift
Background Current
< 8 pA @ 500C
Wander
< 0.5 pA
Iso-thermal drift
< 0.1 pA in 1 hour
Drift under temperature programming from 50o C to 200o C @ 10o C/min
<2 pA
(The drift due to coloumn bleed is much larger.)
For more information and purchase-RELATED QUERRIES, please send us an enquiry here .